Low
Volume Vertical Quarter (Biomekä 2000 Custom
Definition):
Installation
In the Bioworks Edit module select <EDIT>,<Labware>,Click
Modular reservoirs, Select and Copy a Vertical Quarter Modular
reservoir definition ,name it ‘Low Vol. ¼ Vert.’. Edit the
Tracking Height, then Volume. Edit Height and Depth as in figure 1. This
definition will have to be imported into any labbook requiring its use.
Operation:
Positioning is critical to gain accuracy while retrieving from the
micro-trough, Critically align your worksurface. When writing a method the
initial configuration should include this definition on the worksurface.
Although the pull down graphic appears as a Vertical Quarter, it will
operate as intended. Total volume per ea. section is 6000 ul. Volume of
the sub-trough is 10ul. Liquid Level sensing may not perform appropriately
under all conditions however adjustments can and should be made or
standard vertical 1/4s should be used under those conditions. Minor
adjustments to Depth and Height may be required from system to system or
from other reservoir holders like the chillable holder. You will have
to trim the registration nubbs for use with the standard B2K styrene
holder.