TwinivoirX2 low volume
module
Installation
In the Bioworks Edit module select <EDIT>,<Labware>,Click
Modular reservoirs, Select and Copy a Vertical Quarter Modular
reservoir definition ,name it ‘Low Vol. X2 ¼ ’. Edit the
Tracking Height, then Volume. Edit Height and Depth as in figure 1. This
definition will have to be imported into any labbook requiring its use.
Operation:
Positioning is critical to gain accuracy while retrieving from the
micro-trough, Critically align your worksurface. When writing a method the
initial configuration should include this definition on the worksurface.
Although the pull down graphic appears as a Vertical Quarter, it will
operate as intended. Max volume per. Sectionis 4.2mL. Volume of the
sub-trough is 5ul. Liquid Level sensing may not perform appropriately
under all conditions however adjustments can and should be made or
standard vertical 1/4s should be used under those conditions. Minor
adjustments to Depth and Height may be required from system to system or
from other reservoir holders like the chillable holder. You will have to
notch the registration nubbs for use with the standard B2K styrene holder.

Figure 1.